Very low mass isotope data collection with the Nu Vitesse, using single particle multielement ICP-MS for lithium battery electrode materials

Time-of-flight ICP-MS is recognised as a simultaneous elemental analysis technique but the intrinsic qualities of a time-of-flight mass spectrum mean that sensitivity for very low masses is much lower than for higher masses. The use of a multipole device in the reaction cell to remove interferences also introduces a mass bias in the instrument response which is somewhat overcome in the Nu Vitesse by the use of an axial field gradient across the reaction cell, but not completely. The pressurising of the cell and required RF field needed to obtain broad band transmission across the mass range generally means that sensitivities below 23Na are too low to be useful in full mass range spectra. This note will show how very low mass data can be obtained with specific tuning conditions and how this benefits certain applications such as the analysis of battery materials.
 

Simply complete the registration form to the right of this page to download the full note.